The LANSIS 261 features an exclusive implementation of a back illuminated, deep depleted, 2048 x 261 CCD sensor. Deep-depletion combined with high-speed readout and low noise make LANSIS 261 an ideal detector for NIR or Raman spectroscopy.
Key Specs
- Spectroscopy Format
- Deep Cooling
- Sensitivity
- 2048 x 263 array, 15µm pixels
- Liquid cooled to -75 °C
- 95% peak QE, back-illuminated
LANSIS Overview
Highest Reliability
Worry-free, permanent vacuum-seal technology with high performance back-illuminated CCD and the exclusive Teledyne’s eXcelon®technology.
Highest Sensitivity
The LANSIS has the highest average quantum efficiency (QE) from UV to NIR available.
Easiest System Integration
The LANSIS has true plug-n-play convenience. The Teledyne’s ultimate software development kit,(SDK), provides complete control of camera operations. The LANSIS provides a seamless integration to your system.
Specifications
Highlighted
- Resolution
- 2048 x 263
- Megapixels
- 0.5
- Sensor Model
- Back-illuminated AIMO CCD sensor, Grade 1
- Pixel Size
- 15 µm
- Sensor Type
- CCD
- Spectrum
- 200–1100 nm
- Color/Mono
- Mono
- Lens Mount
- Spectroscopy mount
- Data Interface
- USB 3.0
Features
- Triggering Modes
- Start on single trigger, Readout per trigger, Exposure during trigger pulse
Performance
- Dark Current
- 0.02 e⁻/p/s @ -65 °C
- Digitization Bit depth
- 16-bit
- Dynamic Range
- 14,000 : 1
- Full Well Capacity (Qsat)
- 70,000 e⁻ (typical, low gain)
- QE (Mono)
- 95%
- Read Noise
- 5.0 e⁻
- Sensor Material
- Silicon
Physical
- Cooling
- -65 °C (TEC air cooled); -75 °C (liquid cooled)
- Digitization Bit Depth (ADC)
- 16-bit
- Dimensions [W x H x L]
- 151 x 123.5 x 123.5 mm
- Field of View
- 30.72 x 3.96 mm
- Mass
- 3.05 kg, 6.72 lbs
- Operating Humidity
- 0 to 80% (non condensing)
- Operating Temperature
- 5 °C to 30 °C (non-condensing)
- Storage Humidity
- 0 to 90% (non condensing)
- Storage Temperature
- -20 to 60 °C
Recommended System Configuration
- Accessories (Included)
- Power Supply/Cables, MCX to BNC Cables (x3), USB 3.2 Interface Cable, Data Acquisition Software, Certificate of Performance
Compliance & Certifications
- Certification
- FCC Part 15, Subpart B; Class A, CE, UKCA, RoHS 3, PSE, ISO 9001:2015
- Warranty
- 1 year, permanent vacuum guarantee